Instruction No. 154.021.98 IP for the use of “Disposable sterilization indicators IS-120, IS-132, IS-160, IS-180” for monitoring the parameters of the operating modes of steam and air sterilizers

In air sterilizers, control tests are placed at least 5 cm from the walls of the sterilization chamber.
At least one indicator is placed at each point. Indicators IS-160 and IS-180 are attached with an indicator layer outward from the outside of the packages with sterilized products. Fastening is carried out: on packages of polymeric and paper materials and sterilization boxes - with a self-adhesive tape for sterilization, to the tags of sterilization boxes - with a self-adhesive tape for sterilization, a stapler or a paper clip. It is allowed to place indicators inside packages without sterilized products.
The indicators IS-120 and IS-132 are fixed in the same way, placing them both inside the sterilization boxes or packages, and outside.
At the end of the sterilization cycle, the indicators are removed for comparison with the standard. If the color of all indicators matches or darker than the color of the standard, then this indicates compliance with the required sterilization parameters. In this case, a difference in the color of the indicators used between themselves is possible due to the permissible temperature unevenness in the sterilizer. On the indicators IS-120 and IS-132 after processing in the sterilizer, the presence of dark dotted inclusions is allowed, which do not affect the overall color of the indicator.
If the color of the indicator at one or several control points is lighter than the standard, then this indicates non-compliance with the parameters of the sterilization mode. In this case, the entire batch of products is considered unsterilized and is subject to re-sterilization.
When repeating unsatisfactory control results, the operation of the sterilizer is stopped and its technical condition is checked, compliance with the sterilizer loading rules and the correct setting of parameters. The sterilizer is allowed to be used after eliminating the reasons for its unsatisfactory performance and obtaining positive control results.
Used indicators can be pasted into the sterilization accounting journal (form 257 / у) in the columns allocated for this and stored as an archive document for 6 months. For gluing, you can use stationery glue (for example: PVA, glue sticks), except silicate.
<< Previous Next >>
= Skip to textbook content =

Instruction No. 154.021.98 IP for the use of “Disposable sterilization indicators IS-120, IS-132, IS-160, IS-180” for monitoring the parameters of the operating modes of steam and air sterilizers

  1. Guidelines for the control of sterilization using sterilization indicators IS-120, IS-132, IS-160, IS-180 of the Scientific-Production Company "VINAR"
    I APPROVE the Head of the Department of Prevention of the Ministry R.I. KhALITOV Healthcare of the Russian Federation dated 11.06.93 No. 11-8 / 03-54 General Provisions These Guidelines are intended for all health care workers who perform work related to the disinfection and sterilization of medical instruments and medical devices, as well as
  2. Guidelines for monitoring the operation of steam and air sterilizers
    I APPROVE the Head of the Main Epidemiological Department of the USSR Ministry of Health M.I. NARKEVICH dated 02.28.91 No. 15 / 6-5 (Extracts) General Provisions The guidelines are intended for personnel of disinfection and sanitary-epidemiological stations, as well as personnel of medical and preventive institutions (MPIs) that monitor the operation of steam and air sterilizers.
  3. Guidelines for the use of deoxone-4 for disinfection and sterilization
    (Approved by the Head of the State Committee for Sanitary and Epidemiological Supervision of the Russian Federation V.I. Chiburaev dated 06.29.92 No. 11-114 / 3323-1) Designed for employees of disinfection stations, medical departments of territorial centers of the State Sanitary and Epidemiological Surveillance, and treatment and prevention institutions. General information Deoxon-4 is a colorless liquid with a specific smell of vinegar. Contains 5–9% in its composition
  4. Guidelines for the use of deoxone-1 for disinfection and sterilization
    (Approved by the deputy chief of the Main Directorate of Quarantine Infections of the USSR Ministry of Health O. G. Imamaliev from 12.24.80 Ha 28-15 / 6) General Provisions Deoxon-1 is a colorless liquid with a specific smell of vinegar. Contains 5.0-8.0% peracetic acid in its composition. The drug is soluble in water, alcohol and other solvents. Deoxon-1 Solutions Corrode Low Grade Products
  5. Instructions for the disinfection and disinfection of clothes, bedding, shoes and other objects in steam-formalin, steam and combined disinfection chambers and the disinfection of these objects in air disinfection chambers
    (Approved by the Chief State Sanitary Doctor of the USSR on 08.29.77) General Provisions This Instruction contains the basic rules for the disinfection and disinfection of clothes, bedding, shoes and other objects in steam-formalin, steam and combined disinfection chambers and the disinfection of these objects in air disinfection chambers. Disinfection of objects in
  6. Guidelines for the use of the drug "Gigasept FF" for disinfection and sterilization of medical devices (the company "Schulke and Mayer GMBH", Germany)
    (Approved by the Head of the Department of State Sanitary and Epidemiological Supervision of the Russian Federation on 08.31.92) Designed for medical personnel of medical institutions and employees of disinfection stations and territorial centers of the State Sanitary and Epidemiological Surveillance. 1. General information "Gigasept FF" manufactured by the company "Schulke y Mayer" (Germany) is a composition containing as
  7. Sanitary control over the production and use of materials intended for contact with food
    At the food industry enterprises of the corresponding branches of agriculture, public catering and trade systems, various products (utensils, containers, packaging, equipment, implements, consumer goods, etc.) are used in everyday life, made of various materials: metals and alloys, porcelain, faience, glass and a relatively new type of materials (see, below) - polymer.
  8. Steam sterilization
    Steam sterilization is carried out using saturated steam, usually at a temperature of 134 ° C. Staining due to the shedding of chemical indicators A large number of chemical indicators in one sterilized batch can lead to the formation of stains on instruments, especially with direct contact. This applies primarily to products made of silver or with a silver-plated surface.
  9. The effect of physical and chemical parameters of air on human health and hygiene requirements for the air-thermal regime of school premises
    Optimal air conditions are an important factor in maintaining human health and performance. Adverse changes in the air can cause significant disturbances in the body: overheating or hypothermia of the body, hypoxia, the occurrence of infectious and other diseases, decreased performance. In a comprehensive hygienic assessment of air, the following are taken into account: - physical
  10. INSTRUCTIONS FOR USE OF MEDICINES INCLUDING IN THE FIRST AID FIRST AID KIT FOR VESSELS DO NOT HAVE MEDICAL PERSONNEL
    Analgesic and antipyretic drugs 1. Tramal Ampoules in 2 ml. It is used for severe pain: administered intramuscularly or subcutaneously 3-4 times a day. 2. Analgin Tablets of 0.5 g; ampoules of 2 ml of a 50% solution. It is used as an analgesic and antipyretic for pains in muscles, joints, headaches, and febrile conditions. Take orally 1-2 tablets. 2-3 times
Medical portal "MedguideBook" © 2014-2019
info@medicine-guidebook.com